Datex Ohmeda Avance Cs2 User Manual
The Aisys, Avance, Aespire and Aestiva family of anesthesia devices will shut down (stop ventilation) if they run for 49 days consecutively without a reboot. Do not use anesthesia devices for extended time periods without rebooting. If a device has been used for more than 24 hours since the last checkout was completed. Datex-Ohmeda infrared sensor Adjustable low and high alarm limits for EtCO 2 and FiCO 2 Respiration rate (RR) Measurement range: 4 to 60 breaths/min for E-, M-series modules 4 to 100 breaths/min for CARESCAPE modules Detection criteria: 1% variation in CO 2 Adjustable low and high alarm limits for respiration rate; alarm for apnea Patient. Page 1 GE Datex-Ohmeda Cardiocap /5 for Anesthesia ™ Related to software S-XANE 01 User’s Guide 053 7 Conformity according to the Council Directive 93/42/EEC concerning Medical Devices CAUTION: U.S. Federal law restricts this device to sale by or on the order of a licensed medical practitioner.
Page 1
GE Healthcare
Avance CS2
Preoperative Checkout
Refer to the User’s Reference manual for step-by-step instructions.
Every day before your first patient
Check that necessary emergency equipment is available and in good condition.
/summon-night-psp-english-patch-download.html.
Check that the pipeline gas supplies are connected. If equipped with cylinders,
check that there is sufficient reserve capacity and that the cylinder valve is closed.
Atlas translation software download.
Connect scavenging and verify operation.
Check that the breathing circuit is correctly connected, not damaged, and the
breathing system contains sufficient absorbent in the canister.
Turn the System switch On.
Turn the system circuit breaker off and check that mechanical ventilation
continues while the system is running on battery power. After completing the
check, turn the system circuit breaker on. The mains indicator is lit when AC
power is connected.
Set the appropriate controls and alarm limits for the case.
Check that the equipment is not damaged and that components are correctly
attached.
Check vaporizer installation:
• Make sure that the top of each vaporizer is horizontal (not on crooked).
• Make sure each vaporizer is locked and cannot be removed.
• Make sure the alarms and indicators operate correctly (Tec 6 Plus vaporizer).
• Make sure that more than one vaporizer cannot be turned on at the same
time.
• Make sure that the vaporizers are adequately filled.
Perform a Full Test from the Checkout menu.
Check that an adequate reserve O2 supply is available.
Datex Ohmeda Avance Cs2 User Manual User
Check that the ventilator functions correctly:
• Connect a test lung to the patient breathing circuit connection.
• Set the ventilator to VCV mode and the settings to TV to 400 ml, RR to 12, I:E
to 1:2, Tpause to Off, and Pmax to 40 cmH2O.
• Set the gas flow to the minimum settings.
• Start a case.
• Set the Bag/Vent switch to Vent.
• Fill the bellows using O2 flush.
• Check that mechanical ventilation starts. Check that the bellows inflate and
Datex Ohmeda Avance Cs2
deflate. Check that the display shows the correct ventilator data. Check that
Datex Ohmeda Avance Manual
there are no inappropriate alarms.